 |


Comparison of Point-of-Care and Laboratory Glucose
Analysis in Critically Ill Patients
Teresita Lacara, RN, BSN Caroline
Domagtoy, RN, BSN Donna Lickliter, RN Kathy Quattrocchi, RN, BSN
Lydia Snipes, RN Joánne Kuszaj, RN, MSN, CCRN MaryClare
Prasnikar, RN, MSN, CCRN
|
Learning Objectives At the end of
this learning activity, the participant will be able to:
- Discuss use of point-of-care laboratory
testing for glucose analysis in critically ill patients.
- Identify factors that can influence the
accuracy of point-of-care glucose analysis during critical illness.
- Describe how hematocrit values can
influence glucose analysis in critically ill patients.
|
Abnormal blood glucose levels are common in critically
ill patients and increase the risk for complications such as infection,
metabolic problems, and/or cerebral damage.1-5 Frequent monitoring
of blood glucose levels and aggressive management of hyperglycemia can decrease
these complications and mortality.3,6-13 Although laboratory
analysis is the most accurate method for evaluating glucose levels, because of
cost and time delays, bedside point-of-care (POC) testing is often used to
determine glucose levels when frequent monitoring of glucose is important.
Although POC glucose meters were designed to be used with capillary blood
obtained from a fingerstick, clinicians often obtain blood for POC testing from
arterial or central venous pressure (CVP) catheters.
Several investigators14-23 have evaluated
the accuracy of fingerstick POC glucose testing compared with that of
laboratory glucose analysis, but clinical studies24-27 on the
accuracy of using blood from arterial or CVP catheters for POC glucose
determination are limited. None of these studies24-27 included
evaluations of the impact of biochemical derangements, such as altered pH, on
the accuracy of glucose meters when blood from arterial or CVP catheters was
used, and the impact of abnormal hematocrit values25 or poor tissue
perfusion27 was evaluated in only a single study. In earlier
studies14,15,23,28-35 with capillary blood, these conditions
interfered with the accuracy of POC glucose values.
The purpose of this study was to compare POC glucose
meter values of both capillary (fingerstick) and arterial or CVP blood samples
with laboratory glucose values in critically ill patients. In addition, we
examined whether hematocrit, serum level of carbon dioxide, and mean arterial
pressure (MAP) affected the bias of the different sources of blood, because
other investigators14,15,23,25,28,29,31-34 either found or
hypothesized that these physiological variables affected or would affect the
accuracy of measurements obtained with glucose meters.
Materials and Methods
This study was conducted in a 394-bed community-based
hospital in the southeastern region of the United States. Approval was obtained
from the institutions investigational review board before any data were
collected.
Study Design
A method-comparison study design was used to compare
glucose values obtained with a POC device and a clinical laboratory analysis
method. The dependent variables were the differences between glucose values
obtained with the POC testing device (fingerstick and arterial or CVP catheter
specimens) and the values obtained with the clinical laboratory method.
Sample
A convenience sample of critically ill patients had
POC testing done once at the same time a blood sample was collected for glucose
determination via the laboratory method. Inclusion criteria included the
presence of a CVP or an arterial catheter. Sample size was determined a priori
by power analysis (power = 80%, α = .05, effect size = 0.73 for t
test and 0.25 for multiple regression analysis).36 Determinations of
effect size were based on the national standard for minimal acceptable accuracy
for glucose POC devices of 20% or less variation from laboratory values for
glucose concentrations of 75 mg/dL or greater37 (to convert glucose
values to millimoles per liter, multiply by 0.05551).
Procedure
In a standard procedure, blood was obtained from the
CVP or arterial catheter and placed in a separator vacuum test tube for
laboratory glucose testing. The minimum amount of catheter blood discarded
before laboratory glucose sampling was 5 mL, a volume that is at least 5 times
the catheter dead space. Blood for laboratory testing was analyzed by using
standard procedures. Laboratory glucose was determined by using the adapted
hexokinase- glucose-6-phosphate method (Dimension Clinical Chemistry System,
model RxL, Dade Behring Inc, Deerfield, Illinois). Manufacturers
specifications indicate less than 5% mean for the coefficient of variation for
this model of glucose analyzer. 38 Mean time from blood sampling to
laboratory analysis is 45 minutes in our hospital.
After the CVP or arterial catheter blood was obtained
for laboratory testing of glucose, an additional 0.1 mL of blood was obtained
from the catheter for glucose measurement with the POC glucose oxidase
photometry device (SureStep Pro Hospital Meter and SureStep Pro Hospital
Products Test Strips, Johnson & Johnson, Milpitas,
California).38 As specified in the manufacturers directions,
glucose testing was performed by placing a drop of blood on the reagent strip
and then placing the strip into the glucose meter for reading. Capillary blood
was then immediately obtained by lancing a fingertip to produce a drop of
blood, which was then tested for glucose with the POC testing device
(fingerstick POC testing) via the same procedure as that used for catheter
blood. All POC testing was done immediately after blood was obtained from the
catheter or via fingerstick. The same investigator (C.D.) obtained and analyzed
all the POC samples.
Hematocrit, serum carbon dioxide, and MAP values at
the time of therapeutic glucose determination were hand logged onto a data
sheet for each patient. Hematocrit and serum carbon dioxide levels were
analyzed by the hospital clinical laboratory with a Beckman Coulter LH 750
hematology analyzer (Beckman Coulter Inc, Brea, California) and a Dimension RxL
Max w/HM and RMS chemistry analyzer (Dade Behring Inc), respectively, according
to the manufacturers guidelines. MAP was measured directly via a radial
artery catheter connected to a pressure transducer (MX9501T TranStar Patient
Mount Monitoring Kit, Medex, Dublin, Ohio) and a bedside pressure monitor
(Solar 8000, model 415982-005, GE Marquette Medical Systems, Milwaukee,
Wisconsin).
Quality control of the glucose meters was done daily
according to manufacturers directions and included testing both high- and
low-quality control reagents.38 Care was taken to ensure that POC
test strips were from the same lot number and had not expired. Before data
collection, the investigator (C.D.) who would be performing the POC sampling
and analysis was trained in the proper use of the POC glucose
meter.38
Data Analysis
Data were summarized by using descriptive statistics.
Difference scores between both POC glucose values (fingerstick and arterial or
CVP) and laboratory glucose values also were calculated for each patient. Mean
difference scores, or device bias, and limits of agreement between the POC test
values (fingerstick and arterial or CVP) and the laboratory glucose values were
calculated by using the Bland-Altman method.39-41 Also, t
tests were used to determine if differences between the laboratory glucose
value and each of the POC (fingerstick and arterial or CVP) glucose values were
significant. Multiple regression analysis was used to determine if hematocrit,
serum carbon dioxide, and/or MAP accounted for the difference scores between
the laboratory value and the POC glucose values. The level of significance for
all statistical tests was set at P < .05, with a Bonferoni correction
for the multiple t tests.

Results
A total of 49 patients were evaluated. Demographic
data for the sample are presented in Table 1. Mean age was 66.8 (SEM 2.2)
years. Of the 49 patients, 13 had diabetes and 3 had steroid-induced
hyperglycemia. Blood for laboratory testing was obtained from an arterial
catheter in 42 patients and from a CVP catheter in 7. The ranges and mean
values of hematocrit, serum carbon dioxide, and MAP are summarized in Table 2.
Laboratory glucose values ranged from 58 to 265 mg/dL;
fingerstick and catheter POC glucose values ranged from 52 to 281 mg/dL and
from 61 to 263 mg/dL, respectively (Table 3). Glucose values were normally
distributed. Bias (difference) and precision (limits of agreement) were 2.1 and
12.3, with root-mean-square differences (RMSDs) of 12.35 for the fingerstick
POC and laboratory glucose values and 0.6 and 10.6, with RMSD of 10.46 for the
catheter POC and laboratory glucose values (Table 3, Figure 1). Results of the
t tests indicated no significant differences between the laboratory
glucose value and the POC glucose values (fingerstick POC:
t48 = 1.21, P = .23; catheter POC:
t48 = -0.40, P = .69).

Multiple regression analysis indicated that hematocrit
and serum carbon dioxide levels were significant contributors to difference
scores between the laboratory and the catheter POC methods (F3,45 =
8.17, P < .001; Table 4). MAP did not significantly account for the
difference scores. No significant contributors were found for difference scores
between the laboratory and fingerstick POC analysis methods (F3,45 =
2.56, P = .07).
Because the number of samples obtained via the CVP
catheters was so small, the 7 POC blood samples obtained via a CVP catheter
were removed from the study. The bias and precision (differences and limits of
agreement; Table 3, Figure 2) and results of multiple regression analysis
(Table 4) for the arterial POC and laboratory glucose values (n = 42) were
similar to those for the entire sample (N = 49), with the exception of serum
carbon dioxide. Multiple regression analysis for the fingerstick POC method
indicated that serum carbon dioxide level was a significant contributor to the
difference scores between the laboratory and arterial POC analysis methods
(P = .04; Figure 3).

Discussion
Blood obtained from arterial and CVP catheters had
slightly smaller bias and precision values than did blood from a fingerstick
sample. The differences between laboratory and POC glucose values, however,
were not statistically significant. In addition, hematocrit and serum carbon
dioxide levels significantly accounted for the differences between the
laboratory and POC testing values.
Although obtaining blood for POC glucose testing from
arterial or venous catheters instead of using the fingerstick method is common
in critical care units, this practice has been examined in only a few published
studies.24-27 Only 2 of these studies24,25 had sufficient
detail and clarity for adequate interpretation of the results. Small sample
size,24 evaluation of devices no longer used in clinical
practice,26 and inappropriate statistical analysis25,26
limit the generalizability of the studies to clinical practice.
In a study of 50 postoperative cardiac surgical
patients, Maser et al25 found a significant difference between the
mean glucose values obtained with an arterial POC device (Accu-Check II,
Boehringer-Mannheim) and the arterial sample analyzed by using the laboratory
method: 249 (SD 12) mg/dL and 219 (SD 12) mg/dL, respectively. Mean POC values
with fingerstick samples, 210 (SD 12) mg/dL, were lower than the laboratory
mean value. Maser et al hypothesized that the large differences between the
various methods for glucose analysis might have been related to low body
temperature and the effect on the enzymatic reactions on the test strips rather
than to the source of the blood for analysis or to systematic error.
In the study by Maser et al,25 all POC
testing was performed by the same investigator to minimize user error. A major
limitation of the study, which makes interpretation of the data difficult, is
that difference scores between the POC value and the laboratory reference value
were not computed and graphed according to the Bland-Altman
method.39,40 Experts in technology assessment consider the
Bland-Altman method the best way to examine the level of agreement between 2
medical devices.39-41 Correlational methods are inappropriate for
analysis in these types of studies because, with those methods, the degree of
relationship rather than the difference between 2 variables is examined.

Unlike Maser et al,25 Ray et
al24 found no difference between arterial glucose values obtained
with a POC device (One Touch Profile, Lifescan, Johnson & Johnson) and
values obtained with the laboratory reference standard. In that
study,24 blood was obtained from 10 critically ill patients in a
total of 105 sampling periods. The results of our study were similar to those
of Ray et al, but we used a different POC device (SureStep Pro, Lifescan,
Johnson & Johnson).
Hematocrit has long been known to affect the accuracy
of POC glucose analysis.42,43 Despite manufacturers operating
instructions for POC devices that suggest limiting use of the devices to
clinical situations in which hematocrit levels are within a specific range of
values, typically 25% to 55% or 60%, POC devices often are used clinically
without regard for hematocrit level. Similar to our finding, the results of
several studies14,15,25,28,29,31-33 of different glucose meters have
indicated that lower than normal hematocrit values (<30% to <35%) result
in overestimates of laboratory glucose levels when the POC method is used,
whereas hematocrit values higher than normal (>45%) result in underestimates
of laboratory values. Although the mechanism for these differences is not
known, various hypotheses have been proposed to explain the impact of abnormal
hematocrit levels on POC testing: altered viscosity of the blood, prevention of
plasma from reaching the reaction surface of the test strip, change in
diffusion kinetics, and/or increased packed red cell volume and displacement of
plasma volume leading to insufficient plasma volume for accurate
testing.15
In earlier studies14,15,25,28,29,31-33 on
the effect of hematocrit on POC testing, the investigators used fingerstick
(capillary) blood and found a negative bias of glucose values with increasing
levels of hematocrit. We did not find that hematocrit significantly explained
the difference between fingerstick POC glucose values and laboratory values.
This difference in findings may be due to differences in device performance,
because the POC devices in the earlier studies differed from the POC device we
used. Another possible explanation is that hematocrit ranges in our study were
more homogeneous than were the ranges in the other studies. Although hematocrit
values ranged from 20% to 53% in our study, mean hematocrit was 32% with a
small SEM (0.8%). Few of our subjects had hematocrit values outside the
manufacturers recommended range for optimal device functioning (25%-60%).
Another reason for the different results might be the
changes made in POC devices used today to improve the accuracy of measurement
of capillary blood glucose levels in patients with abnormal hematocrit levels.
This last reason also would explain why hematocrit levels significantly
accounted for differences between POC glucose values obtained with CVP or
arterial blood and the laboratory value in our study but not for the
differences between POC values with capillary (fingerstick) blood and the
laboratory value. Any improvements in device performance made to correct the
effect of hematocrit on glucose determinations when the POC device is used with
capillary samples could then cause an overcorrection if the source of blood
used for POC testing is not capillary blood.
In addition to the adverse effect of hematocrit levels
on the accuracy of POC glucose values, several other factors can interfere with
the accuracy of different models of POC glucose meters44: creatinine
levels,45,46 bilirubin levels,30 high arterial oxygen
levels,45,46 shocklike states,23,34,35,47 and altered
pH.31,34 On the basis of anecdotal experience with discrepancies
between POC and laboratory glucose values in individual patients with shock
states, we decided to evaluate MAP and serum carbon dioxide level, because
arterial pH values are not routinely available in our patients. Serum carbon
dioxide levels vary in relationship to alterations in blood pH, so we
hypothesized that serum carbon dioxide would be associated with differences
between POC and laboratory glucose. In all but one of the multiple regression
analyses we performed (fingerstick POC glucose values minus laboratory glucose
values), serum carbon dioxide significantly accounted for the POC and
laboratory difference scores. This finding is similar to, and consistent with,
findings in previous studies31,34 on the effect of pH on the
accuracy of POC glucose values, because serum levels of carbon dioxide depend
on serum pH.
Unlike the findings in previous
studies23,34,35,47 of shocklike states, MAP did not significantly
explain the difference scores between POC and laboratory glucose values in our
study. Because most of our subjects had MAP greater than 60 mm Hg, and thus did
not have shocklike states, our findings are not surprising. Future studies
should be done in patients with shock to determine if MAP is a significant
contributor to POC accuracy.
Clinical Implications
The lack of differences between glucose values
obtained via a laboratory analysis and values obtained with the POC device,
whether capillary or catheter blood was used, supports the common practice in
critical care units of using catheter rather than fingerstick blood for POC
testing. With the advent of aggressive glucose management protocols to decrease
infection risks in critically ill patients,3,6-13 blood samples are
needed frequently, and use of catheter blood avoids painful needlesticks to
obtain capillary blood.
One limitation of our findings is that all of the POC
tests were done by the same investigator. This practice, although ideal for
research control, is not typical of the way POC testing is usually done.
Significant user error can be introduced with multiple users and can make the
difference between POC and laboratory values in actual practice situations more
pronounced than the differences in our study. Having multiple care providers do
the POC testing in a replication of our study would be prudent to determine the
impact of the number of testers on the accuracy of POC glucose values. In the
other study25 that showed no difference between POC and laboratory
glucose values, a single provider also did all the POC tests.
Because of the range of difference scores and the
width of the 95% CIs (the ±2 SD lines) of the Bland-Altman graphs
(Figures 1 and 2) in our study, we recommend caution in using any individual
POC glucose value as a basis for adjusting insulin doses when tight glucose
management protocols are being used. Many of the patients in our study had
glucose differences of at least 10 mg/dL between the POC and laboratory
methods, which may be a large enough difference from the true glucose value to
change management decisions when treatment protocols call for narrow ranges for
glucose levels.
As shown in other
studies,14,15,25,28,29,31-33 abnormal hematocrit levels may
influence the accuracy of a POC device. In patients with abnormally low
hematocrit levels, periodically validating the POC test result by comparing it
with a laboratory glucose level would be a prudent approach to avoid situations
in which the POC value is an overestimation of the true glucose value.
Conclusions
Our findings validated the practice in critical care
units of obtaining blood for POC testing from arterial or venous catheters
rather than from a fingerstick source. The bias and precision of glucose POC
testing with a fingerstick sample was slightly higher than the bias and
precision with samples obtained from an arterial or a CVP catheter, but the
differences were not statistically significant. In addition, for hematocrit and
serum carbon dioxide levels below normal ranges, POC values with arterial blood
tended to be overestimates of laboratory glucose values. In individual
situations, for patients with abnormal hematocrit or carbon dioxide levels,
care should be taken to verify the accuracy of POC glucose testing by comparing
the POC value with the laboratory value of a sample obtained at the same time
as the POC sample. Further study is needed to determine if patients with
abnormally low MAP have greater discrepancies between POC and laboratory
glucose values.

Acknowledgments
Special thanks to Marianne Chulay, RN, DNSc, FAAN, for
assistance with study design, data analysis, and manuscript preparation.
Financial Disclosures
None reported.
References
- Umpierrez GE, Isaacs SD, Bazargan N, You X, Thaler
LM, Kitabchi AE. Hyperglycemia: an independent marker of in-hospital mortality
in patients with undiagnosed diabetes. J Clin Endocrinol Metab.
2002;87:978-982.
- Van den Berghe G, Wouters P, Weekers F, et al.
Intensive insulin therapy in critically ill patients. N Engl J
Med. 2001;345:1359-1367.
- Pomposelli JJ, Baxter JK III, Babineau TJ, et al.
Early postoperative glucose control predicts nosocomial infection rate in
diabetic patients. J Parenter Enteral Nutr. 1998;22:77-81.
- Capes SE, Hunt D, Malmberg K, Gerstein HC. Stress
hyperglycaemia and increased risk of death after myocardial infarction in
patients with and without diabetes: a systematic overview. Lancet.
2000;355:773-778.
- Pierre EJ, Barrow RE, Hawkins HK, et al. Effects of
insulin on wound healing. J Trauma. 1998;44:342-345.
- Dagogo-Jack S, Alberti KG. Management of diabetes
mellitus in surgical patients. Diabetes Spectr. 2002;15:44-48.
- Dellinger EP. Preventing surgical site infections:
the importance of timing and glucose control. Infect Control Hosp
Epidemiol. 2001;22:604-606.
- Jacober SJ, Sowers JR. An update of perioperative
management of diabetes. Arch Intern Med. 1999;159:2405-2411.
- Malmberg K, Ryden L, Efendic S, et al. Randomized
trial of insulin-glucose infusion followed by subcutaneous insulin treatment in
diabetic patients with acute myocardial infarction (DIGAMI study): effects on
mortality at 1 year. J Am Coll Cardiol. 1995;26:57-65.
- Furnary AP, Gao G, Grunkemeier GL, et al.
Continuous insulin infusion reduces mortality in patients with diabetes
undergoing coronary artery bypass grafting. J Thorac Cardiovasc
Surg. 2003;125:1007-1021.
- Van den Berghe G, Wouters PJ, Bouillon R, et al.
Outcome benefit of intensive insulin therapy in the critically ill: insulin
dose versus glycemic control. Crit Care Med. 2003;31:359-366.
- Clement S, Braithwaite SS, Magee MF, et al.
Management of diabetes and hyperglycemia in hospitals [published corrections in
Diabetes Care. 2004;27:836, 1255]. Diabetes Care. 2004;27:553-591.
- Krinsley JS. Effect of an intensive glucose
management protocol on the mortality of critically ill adult patients
[published correction in Mayo Clin Proc. 2005;80:1101]. Mayo Clin
Proc. 2004;79:992-1000.
- Puntmann I, Wosniok W, Haeckel R. Comparison of
several point-of-care testing (POCT) glucometers with an established laboratory
procedure for the diagnosis of type 2 diabetes using the discordance rate: a
new statistical approach. Clin Chem Lab Med. 2003;41;809-820.
- Louie RF, Tang Z, Sutton DV, Lee JH, Kost GJ.
Point-of-care glucose testing: effects of critical care variables, influence of
reference instruments, and a modular glucose meter design. Arch Pathol
Lab Med. 2000;124;257-266.
- Tang Z, Du X, Louie RF, Kost GJ. Effects of pH on
glucose measurements with handheld glucose meters and portable glucose analyzer
for point-of-care testing. Arch Pathol Lab Med. 2000;124:577-582.
- Velazquez Medina D, Climent C. Comparison of
outpatient point of care glucose testing vs venous glucose in the clinical
laboratory. P R Health Sci. 2003;22;385-389.
- Poirier JY, Le Prieur N, Campion L, Guilhem I,
Allannic H, Maugendre D. Clinical and statistical evaluation of self-monitoring
blood glucose meters. Diabetes Care. 1998;21:1919-1924.
- Fanghanel G, Sanchez-Reyes L, Morales M, et al.
Comparative accuracy of glucose monitors. Arch Med Res.
1998;29:325-329.
- Bohme P, Floriot M, Sirveaux MA, et al. Evolution
of analytical performance in portable glucose meters in the last decade.
Diabetes Care. 2003;26:1170-1175.
- Harding K. A comparison of four glucose monitors in
a hospital medical surgical setting. Clin Nurse Spec.
1993;7:13-16.
- Aboezz R, Miller D. Accuracy of portable blood
glucose monitors. J Am Pharm Assoc (Wash DC). 2005;45:514-516.
- Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI,
Reddy S. Fingerstick glucose determination in shock. Ann Intern
Med. 1991;114:1020-1024.
- Ray JG, Hamielec C, Mastracci T. Pilot study of the
accuracy of bedside glucometry in the intensive care unit. Crit Care
Med. 2001;29:2205-2207.
- Maser RE, Butler MA, DeCherney GS. Use of arterial
blood with bedside glucose reflectance meters in an intensive care unit: are
they accurate? Crit Care Med. 1994:22;595-599.
- Pressly KB, Batteiger TH, Barnett DZ, Woodie ME.
Use of arterial blood for glucose measurement by reflectance. Nurs
Res. 1990;39:371-373.
- Kanji S, Buffie J, Hutton B, et al. Reliability of
point-of-care testing for glucose measurement in critically ill adults.
Crit Care Med. 2005;33:2778-2785.
- Phillipou G, Seaborn CJ, Hooper J, Phillips PJ.
Capillary blood glucose measurements in hospital inpatients using portable
glucose meters. Aust N Z J Med. 1993;23:667-671.
- Smith EA, Kilpatrick ES. Intra-operative blood
glucose measurements: the effect of haematocrit on glucose test strips.
Anaesthesia. 1994;49:129-132.
- Jain R, Myers TF, Kahn SE, Zeller WP. How accurate
is glucose analysis in the presence of multiple interfering substances in the
neonate? J Clin Lab Anal. 1996;10:13-16.
- Tang Z, Lee JH, Louie RF, Kost GJ. Effects of
different hematocrit levels on glucose measurements with handheld meters for
point-of-care testing. Arch Pathol Lab Med. 2000;124:1135-1140.
- Wiener K. The effect of haematocrit on reagent
strip tests for glucose. Diabet Med. 1991;8:172-175.
- Barreau PB, Buttery JE. Effect of hematocrit
concentration on blood glucose value determined on Glucometer II.
Diabetes Care. 1988;11:116-118.
- Kilpatrick ES, Rumley AG, Smith EA. Variations in
sample pH and po2 affect ExacTech meter glucose measurements. Diabet
Med. 1994;11:506-509.
- Sylvain HF, Pokorny ME, English SM, et al. Accuracy
of fingerstick glucose values in shock patients. Am J Crit Care.
1995;4:44-48.
- Cohen J. Statistical Power Analysis for the
Behavioral Sciences. Rev ed. New York, NY: Academic Press; 1977:313.
- International Organization for Standardization. In
Vitro Diagnostic Test Systems Requirements for Blood-Glucose Monitoring Systems
for Self-testing in Managing Diabetes Mellitus. Geneva, Switzerland:
International Organization for Standardization; 2003. Publication 15197.
- SureStep Pro Hospital Meter Product Guidelines.
Milpitas, CA: Johnson & Johnson Inc.
- Bland JM, Altman DG. Statistical methods for
assessing agreement between two methods of clinical measurement.
Lancet. 1986;1:307-310.
- Bland JM, Altman DG. Measuring agreement in method
comparison studies. Stat Methods Med Res. 2000;8:135-160.
- Chatburn R. Evaluation of instrument error and
method of agreement. AANA J. 1996;64:261-268.
- Consensus statement on self-monitoring of blood
glucose. Diabetes Care. 1987;10:95-99.
- American Diabetes Association. Clinical practice
recommendations 1996. Diabetes Care. 1996;19(suppl 1):S1-S118.
- Using blood glucose meters: minimizing errors,
maximizing accuracy. Health Devices. 2004;33:251-256.
- Kurahashi K, Maruta H, Usuda Y, Ohtsuka M.
Influence of blood sample oxygen tension on blood glucose concentration
measured using an enzyme-electrode method. Crit Care Med.
1997;25:231-235.
- Zaloga GP. Beware of errors in blood glucose
measurements [editorial]. Crit Care Med. 1997;25:212.
- Kulkarni A, Saxena M, Price G, OLeary MJ,
Jacques T, Myburgh JA. Analysis of blood glucose measurements using capillary
and arterial blood samples in intensive care patients. Intensive Care
Med. 2005;31:142-145.
Reprinted from American Journal of Critical
Care, July 2007, pp 336-346 © 2007, AACN.
TOP |
 |